1) Elements: Cd,As, Pb
2) Accuracy: ≤15%
3) LOD: Cd≤0.05mg/kg, As & Pb≤0.08mg/kg
4) Precision: given one same analyzer with 6 constant detection by a same operator for a same sample in a same lab, Cd RSD≤5%,As & Pb RSD≤10%.
5) Inter-instrument agreement: given detection by different devices for a same sample, there is a chance of ≤5% that the RSD exceeds 20%
6) Radiation: no radiation leak which meets the state radiation measurement standard.
7) Detector(imported): FAST-SDD, all-in-one design ,with energy resolution≤129ev
8) Integrated X-ray generator: side-window tube, target material in Tungsten, electrical force≥60KV, electrical current≥1mA, power: 12W.
9) Two modes: Fast detection and defined-quantitative detection.
10) Analytical results will be presented automatically in file with its form customized based on your need.
11) Recyclable sample cups